What happens when the newly bought laboratory water cooling system triggers alarm in the first start?

Some users bought the new laboratory water cooling systems and when they first started the chiller,the alarm was triggered. Well,it is not a big problem and that’s common for the new water cooling system.

water cooling system

Some users bought the new laboratory water cooling systems and when they first started the chiller, the alarm was triggered. Well, it is not a big problem and that’s common for the new water cooling system. Users can deal with this alarm by following the below advice:

1.First, turn off the water cooling system and use a pipe to short connect the water inlet and water outlet. Then turn on the chiller to see if the alarm continues;

1.1If the alarm disappears, it is possible that there is blockage in the external water channel or the pipe is bent;

1.2If the alarm continues, it is possible that there is blockage in the internal water channel or the water pump;

If the above conditions are ruled out and the alarm continues, it is likely that the components are faulty. But this is pretty rare, for all of S&A Teyu water cooling systems are under strict quality control before delivery. 

After 18-year development, we establish rigorous product quality system and provide well-established after-sales service. We offer more than 90 standard water chiller models and 120 water chiller models for customization. With cooling capacity ranging from 0.6KW to 30KW, our water chillers are applicable to cool different laser sources,laser processing machines, CNC machines, medical instruments, laboratory equipment and so on.

laboratory water chiller

We're here for you when you need us.

Please complete the form to contact us, and we'll be happy to help you.

Home       Products          SGS & UL Chiller       Cooling Solution       Company       Resource       Sustainability
Copyright © 2025 TEYU S&A Chiller | Sitemap    Privacy policy
Contact us
email
Contact customer service
Contact us
email
cancel
Customer service
detect